Fer Strumenti Mod 5185 Oxygen analyzer uses a fully
stabilized Zirconia cell. Long experience with high temperature oxygen
analyzers ensures total confidence in this model which incorporates a heating
element to raise the cell to the working temperature.
To obtain the correct air/fuel ratio in a combustion process it is essential
to measure the content of oxygen present in the flue gas.
Such measurement is possible with the 5185 Analyzer which measures the
percentage of O2 directly in the stack or ducts; without the need for gas
extraction
PROBE
The probe has been studied to enable the
cheap oxygen measuring in a flue gas
even in small dimensions boilers and furnaces. Fully stainless steel for the
part in contact with the combustion products, this model has the same working
characteristics of the models studied for bigger boilers.
No reference air is needed for this model due to short insertion length direct
on the stack (in situ).
TECHNICAL SPECIFICATIONS
-
Accuracy : ±
1% of theoretical value or 0,1 O2
(whichever is the greater) if used with Fer
Strumenti transmitters.
-
Stability : within 1%
-
Operating temperature :
-
400 °C max. for the part inside the process
-
150 °C
max. for the head of the probe.
-
Thermocouple : type K
-
Heater : 230 V or 110 V - 200 V A max.
Process connection : male thread
1/2 GAS
-
Dimensions : Ø of wet part mm. 43,
length mm. 200
-
Weight : kg 4,5
|