|
Analyzers and systems for flue and combustible gas |
|
Infrared analyser Mod. Enox DPB for very low ranges |
||
ImtroductionThe
analyzer is housed in a steel 19” rack 3 vertical unit high.
On the rear panel are
located the connector for digital and analogical input and output, the
filtered power connect with switch and fuses and the fittings for sample
inlet and outlet. Inside the housing is located a low flow detector and
optionally a diaphragm pump and an AISI solenoid valve for automatic zero
calibration and an electrochemical cell for the O2 simultaneously measure
in the sample. In the front panel are present a 16 keys keyboard, a fast to be check and replace fine filter with the big back lighted LCD display ApplicationThe analyzer can be used for gas measuring in a wild range of industrial applications:
Principle of operation
A wide band infrared
radiation is emitted by a non metallic, high stability source. For
each measured gas this radiation is brought alternatively through an
interferential filter and a filter with a glass cell (GFC) filled with
high partial pressure of the gas to be analyzed. A
suitable optical system drives the IR radiation inside the analysis
chamber and then to the detector which receives and amplifies
alternatively the two signals: one is the measure, the other is the
reference. The gas concentration is proportional to the difference of the
two signals. Gases
which may have a cross sensitivity with the measured gas, generate the
same variation of the measure and reference signals. Therefore the measure
is not affected. Technical description The
Enox multigas analyzer is an industrial photometer based on the non
dispersive photometry in the infrared, for the contemporary measure of
several gases. The
main feature of this enhanced analyzer model DPB (double path beam) is to
have a long optical path to obtain to very high sensitivity to guarantee
the declared performance even on the very low ranges requested from the
lower and lower emissions limit
imposed. The
mainly pulse to follow and reach this solution has been driven from the
necessity to develop an NDIR NO (nitrogen oxide) analyzer with performance
comparable with the expensive ones normally used where the emission limits
are about some tens vpm.
The
technology which has been used is based on correlation filters (GCF) and
optical non dispersive absorption (DOAS). A high stability sensor, working
at a very low temperature (-35 °C) and the GFC technology ensure a nearly
fully immunity to cross sensitivity, high stability and sensitivity. The
measuring principle, the optical bench and the automatic and accurate
compensation of the variations of room temperature enable to avoid the
expensive and complicated automatic calibrations, which are always
possible. A
large back lighted graphic display continuously supplies the value of the
measured gases (including a large bar graph), the alarms, the service
requirements, the faults, separated for each gas and the alarms from the
sampling system. All alarms have the acknowledgement routine. All anomaly
messages are reported on the display for the immediate knowledge of what
has gone wrong. Relays contacts enable all the diagnostic to be
retransmitted. This
instrument has been drawn to be user friendly and reduce as much as
possible the costs of the analysis system in which it will be used. The
analyzer can manage a multiplexing CEM up to four emission point and can
be directly connected to a P.C. provided with a special software, named
DAS-DAC, for data acquisition monitoring and reporting..
|
||
|
Technical specifications |
||
|
Measured components |
Lowest ranges: 1. CO 0…10 ppm 2. CO2 0…10 ppm 3. NO 0…50 ppm 4. SO2 0…400 ppm 5. CH4 0…100 ppm 6. O2 0…25% in volume Highest ranges: 100% or saturation. |
|
Measure units |
vpm, mg/m3, mg/Nm3, % |
|
Display |
240X128 pixel, graphic, back lighted LCD, tunable contrast |
|
Keyboard |
16 keys membrane |
|
Analogical Output |
N. 4 4-20 mA linear isolated. Max load 500 ohm. |
|
Digital Output |
RS 232, RS 485, Modbus |
|
Flow control on sample |
Continuous on flow lower than 0.5 l/min |
|
Digital Output |
N. 2 contacts settable alarms on each measure, except oxygen N. 1 contact service for each measure, except oxygen N. 1 contact fault for each measure, except oxygen, and flow fault N. 1 contact for calibration N. 2 contacts for solenoid valve, in case of external calibration (24 Vcc 50 mA) |
|
Digital Input (12 Vcc 100 mA) |
N. 1 Distant calibration N. 1 sampling system fault N. 4 stack currently on measure (for scansions) N. 4 low cylinder pressure |
|
Warm up time |
’15 45’ best performance |
|
Zero calibration. |
Automatic with ambient air or nitrogen. Tunable frequency and duration |
|
Span calibration |
Automatic not necessary. Tunable frequency and duration |
|
Drift |
Negligible with automatic zero calibration: · < 2% of lowest range without automatic zero calibration · Ambient temperature: negligible continuously compensated · Atmospheric pressure: settable · zero: none · span about 1% of measured value for 1% of atmospheric pressure change |
|
Response time (T90) |
Depending on mobile average set |
|
Noise |
</= 1% of lowest range |
|
Mobile average |
Tunable between 0 ….600” |
|
Ambient temperature |
For measure +5…+35°C For stocking and transport -10…+55°C |
|
Moisture |
<90% RH not condensing |
|
Panel filter |
Retention: 1 micron |
|
Sample gas status |
pressure 20…80 mbar flow 30…180 Nl/h temperature +5…+50°C dew point at least 5°C under ambient temperature |
|
Protection |
IP20 |
|
Power |
115-230 V AC +/-10% about 70 VA |
|
Dimensions |
450X132x380 mm |
|
Weight |
kg. 12 |